Operation
GMS810-/GMS811-FIDOR · Operating Instructions · 8013911 V1-0 · © SICK AG 55
Subject to change without notice
5.3 Starting important operating sequences
5.3.1 Check and adjustment with test gas
Step Action Reference
1 Connect zero and reference gas.
p. 23, §3.4
2 Switch device to “Maintenance”: Menu Maintenance/Mainte-
nance mode/Maintenance mode.
p. 51, §5.2.8.2
3 Set reference gas parameters: Menu Parameter/Reference gas.
p. 46, §5.2.6.3
4 If necessary, set gas timing parameters: Menu Parameter/Gas tim-
ing.
p. 47, §5.2.6.7
5 Start zero and reference point adjustment: Menu Adjustment/Zero
+ reference
p. 48, §5.2.7.2