5.2.5.5 Telediagnostic
Menu: FIDOR/Diagnostic/Telediagnostic
This menu shows an internal diagnosis (for service purposes).
Temperature Flame Actual flame temperature
Detector Actual temperature of analysis chamber
Line Actual temperature of heated line
Catalyst Actual temperature of catalyst
Pt100-4 Not used
pAA Actual temperature of pA amplifier
Electronics Actual temperature of electronics board
Pressure P1 Actual pressure detector
P2 Actual sample gas inlet pressure
P4 Actual combustion air pressure
P5 Actual fuel gas pressure
P3 Actual control air pressure
P3-P2 Pressure difference P3-P2
Voltage Suction Actual suction voltage
3.3V Actual value of 3.3 V voltage supply
5V Actual value of 5 V voltage supply
24V Actual value of 24 V voltage supply
Current Ignition Actual ignition current of glow filament in A
Power pAA Heating output of pA-amplifier
Detector Heating output of detector
Line Heating output of heated line
Catalyst Heating output of internal catalyst
Electric Power of electronics
Total Total power
Maximum Maximum allowed power
Menu item Significance
Maintenance
request
Short form of message (for a detailed description, see menu further on).
Failures Short form of message (for a detailed description, see menu further on).
Monitoring Status tracking of start process.
Start state State saved after power-On.
Maint.req. xx Actual pending maintenance requests.
Failures xx Actual pending causes for failure.
Language Language of telediagnostic error messages (German or English).
Menu item Submenu Significance